If the PDM has some anomalies during the course of use, the trouble can be basically shot by checking the problems against the following table. If you meet any problems not covered in the following table, you may contact our technical service department to deal with your problems.
PDM failure analysis and troubleshooting table
Anomalies
|
Possible Cause
|
Diagnosis and troubleshooting methods
|
The riser pressure gauge index suddenly elevated
|
Motor drag
|
Lift the drill 0.5-1m away from the bottom of the well; if the circulating pressure is correct, slowly increase the bit weight.
|
The drive shaft of the motor is stuck
|
Lift the drill 0.5-1m away from the bottom of the well; if the circulating pressure is too high, pull out the drill for checking and change the drill.
|
Standpipe pressure gauge index increased, but not with the drilling depth increases growth of normal pressure drop
|
Ice drill’s melting water eyes or the drill’s internal water eye was blocked
|
Lift the drill off the bottom of the well; if the circulating pressure is still higher than the normal value, change the cyclic flow or move the drill up and down; if it not effective, pull out the drill for check and change the drilling bit.
|
Drill bit wear
|
Continue drilling if there is not any footage, pull out the drill for check and replace the drill bit.
|
Stratigraphic changes
|
Lift the drill off the bottom of the well; if the circulating pressure is normal, continue drilling.
|
The pressure gauges index decreased slowly
|
Drill pipe damage
|
Lift the drill off the bottom of the well; if the circulating pressure is lower than the normal value, pull out the drill for check.
|
Change of circulating pressure loss
|
Check whether the drilling fluid flow is changed
|
No footage
|
Stratigraphic changes
|
Appropriately change the flow and bit weight within the allowable range
|
The bypass valve is in the “open position”
|
Lift the drill off the bottom of the well; if the circulating pressure is lower than the normal value, start and stop the mud pump respectively twice. If it is not effective, pull out the drill to check the bypass valve.
|
Motor drag
|
Lift the drill 0.5-1m away from the bottom of the well; if the circulating pressure is correct, slowly increase the bit weight
|
Cardan shaft damage
|
It is accompanied by pressure fluctuations and if the PDM is lifted, the pressure fluctuation range will be reduced; pull out the drill for check and replace the drill.
|
Drill bit wear
|
If the circulating pressure is normal, pull out the drill for check and replace the drill bit.
|